Enables on-site check and calibration of acceleration pickup, vibration measurement system and vibration level meter.
Feature
・Low cost and provide accurate sine wave vibration without distortion.
・Calibration with acceleration, velocity and displacement(10m/s²,10mm/s,10μm) are possible.
・Small size(φ68 x 200mm) and light weight (approx.600g) make it easier to bring the calibrator to measurement sites.
・Grip designed for easy grasp.
・Suitably shaped to prevent from falling.
・Have built-in pickup for reference level control and temperature sensor.
・Calibration of pickup up to 70g in weight is possible.
・Carrying case included.
Example of usage
Firmly fix the acceleration pickup with the screw attached to the product or by using thin double-stick tape.
Push the push button on the side of the body to start the operation.
Adjust the measurement equipment so that correct reference value is indicated.
Vibration Frequency | 159.2Hz ±1%(-10℃~+55℃) | |
---|---|---|
Vibration Acceleration | 10m/s² (rms) 14.14m/s²(peak) | ±3%(+10℃~+40℃) ±5%(-10℃~+55℃) |
Vibration Velocity | 10m/s(rms) 14.14m/s(peak) | ±4%(+10℃~+40℃) ±6%(-10℃~+55℃) |
Vibration Displacement | 10μm(rms) 14.14μm(peak) | ±5%(+10℃~+40℃) ±7%(-10℃~+55℃) |
Lateral Vibration Amplitude | Less than 5% of the amplitude in main axis. | |
Limit of the mass of the acceleration pickup to be calibrated | 70g or less. | |
Fluctuation of vibration due to difference in mass of the acceleration pickup to be calibrated | ±3%(acceleration +10℃~+40℃) | |
Total Harmonic Distortion | 3% or less (Mass of the acceleration pickup to be calibrated 20~60g) 5% or less (Mass of the acceleration pickup to be calibrated 10~70g) |
|
Overload sensing function | Calibration is automatically stopped if the mass exceeds 100g. | |
Operation Mode | Operation is automatically stopped.(after approx.2 minutes) | |
Battery | Size AA battery 4ea. | |
Operating Temperature and Humidity | -10℃~+55℃, 90%RH or less. | |
Dimension and weight | φ68mm×200(L)mm・ Approx.900g(including battery weight) |
1G Calibrator TYPE 2101 | 1ea |
---|---|
M6/M3 Adapter Screw | 1ea |
M6 Flat Screw | 1ea |
Carrying Case | 1ea |
Size AA battery | 4ea |
Instruction Manual | 1ea |